Experimental Stations
Home › Facilities › Experimental Stations
Detectors
ADSC Quantum-4
- Active Area
188 x 188 mm - Resolution
2304 x 2304 pixels
81.6µm pixel size - Read-out (16 bit)
3-11 seconds (mode dependent) - Dynamic range
65,535 - Dark Current
0.03/pixel/sec - Def.site file
14-BM-D ADSC Q4 - download
(PC users, right click the link and choose "Save Target As...")
ADSC Q315
- Active Area
315 x 315 mm - Resolution
6144 x 6144
51µm pixel size - Read-out (16 bit)
1 second - Dynamic range
65,535 - Dark Current
0.015 e/pixel/sec - Def.site file
14-BM-C ADSC 315 - download
(PC users, right click the link and choose "Save Target As...")
MAR345
- Active Area
180 - 345mm (diameter, mode dependent) - Resolution
0.1 - 0.15mm (mode dependent) - Read-out
120 seconds
MAR Readout Modes
| mode | pixel (mm) | readout diameter (mm) | image size (pixels) | file extension |
|---|---|---|---|---|
| 0 | 0.15 | 345 | 2300*2300 | .mar2300 |
| 1 | 0.15 | 300 | 2000*2000 | .mar2000 |
| 2 | 0.15 | 240 | 1600*1600 | .mar1600 |
| 3 | 0.15 | 180 | 1200*1200 | .mar1200 |
| 4 | 0.10 | 345 | 3450*3450 | .mar3450 |
| 5 | 0.10 | 300 | 3000*3000 | .mar3000 |
| 6 | 0.10 | 240 | 2400*2400 | .mar2400 |
| 7 | 0.10 | 180 | 1800*1800 | .mar1800 |
mar CCD 165
- Active Area
162 mm diameter - Resolution
2048 x 2048 pixels
79µm pixel size - Read-out (16 bit)
3 seconds - Def.site file
14-ID-B MarCCD 165 - download
(PC users, right click the link and choose "Save Target As...")
Kappa Diffractometer
A Kappa geometry diffractometer is available in all three BioCARS stations: 14-BM-C, 14-BM-D and 14-ID-B.
The angular range of omega and kappa motions is limited by possible collisions with the collimator box, beam stop, detector and detector mount.
For detector distances over 100mm, it is safe to rotate omega between -90° and +90° when kappa is at 0° (usual data collection geometry). For shorter distances, inclined detector geometry or inclined kappa, please contact the BioCARS staff.
Samples can be mounted with horizontal spindle (omega=0°, kappa=0°) or inclined spindle (omega=0°, kappa=+60°). Fluorescence scans are done at omega=0° and kappa=-60°.
Phi rotation is used for centering of the sample and omega rotation for oscillation data collection.
Travel ranges are:
X-axis: -5 to +5 mm
Y-axis: -5 to +5 mm
Z-axis: -12 to +12 mm
Pins (with loops or capillaries) or magnetic bases are mounted on a drill chuck. Two chuck sizes are available: the large one accommodates 1 to 5 mm pin diameters and the small one accommodates 1 to 3 mm diameters.
The Z-axis travel range accommodates 15-35 mm cryo loop pin lengths (measured from the magnetic base) when the small chuck is used and 25-35 mm pin lengths when the large chuck is used.