The 14-BM-D station is a bending magnet station used for MAD phasing.
Optics
- Si (111) double-crystal monochromator
- bent-flat Si-mirror (Rh coated)
Beam Characteristics
- Energy range:
6.8 - 17.5 keV - Bandpass:
ΔE/E: 1.3 x 10-4 - Spot size, horizontal x vertical (FWHM):
0.5 x 0.8 mm2 (can be reduced using slits)
Station
- Sample positioning:
kappa diffractometer; inclined detector geometry supported
goniometer head with remote xyz-translation
B&W and color sample alignment camera - Sample to detector distance:
70 to 800 mm adjustable - Sample temperature control:
liquid nitrogen sample cooler - Detectors:
ADSC Quantum-4